ASAP 2020 Confirm
Equilibration
Equilibration interval
table
Minimum equilibration
delay at P/Po> = 0.995
Backfill
4-28
Displays the Equilibration dialog.
Use this table to enter pressure ranges and intervals between
them. Long equilibration intervals tend to lengthen analyses,
however, they do improve data integrity. Short equilibration
intervals produce a faster analysis but may reduce the
accuracy of data.
The value entered in this field determines the minimum
number of seconds required before equilibration can occur
for a relative pressure greater than or equal to 0.995.
This field does not display if you select Absolute pressure
dosing on the Analysis conditions dialog.
Displays the Sample Backfill Options dialog, allowing you to
select options for backfilling the sample tube at the start and
end of an analysis.
New
Does not
display when
Absolute
dosing is
selected.