Quotes, Orders, and Customer Satisfaction Customer Service Phone: 1-770-662-3636 Orders@Micromeritics.com --------------- U.S. Instrument Service Department U.S. Service Department Phone: 1-770-662-3666 Service.Helpdesk@Micromeritics.com --------------- Micromeritics Scientific Services Customer Support Portal http://techsupport.micromeritics.com/portal TriStar II 3020 Confirm Installation Instructions and Checklist 302-42870-04 (Rev A) — Sep 2017...
CAUTION - Cautions contain information to help prevent actions that may damage the analyzer or components. WARNING - Warnings contain information to help prevent actions that may cause per- sonal injury. TriStar II 3020 Confirm Installation Instructions and Checklist 302-42870-04 (Rev A) — Sep 2017...
Connect a Regulator and a Gas Line to a Gas Cylinder 4 - 6 Connect Gas to Analyzer 4 - 8 Clean and Verify the Gas Line 4 - 9 Specify Gas Ports 4 - 12 TriStar II 3020 Confirm Installation Instructions and Checklist 302-42870-04 (Rev A) — Sep 2017...
Page 6
A - 1 Carbon Reference Material Run Conditions A - 4 B Worksheets B - 1 Sample Data Worksheet B - 2 User Information Worksheet B - 3 TriStar II 3020 Confirm Installation Instructions and Checklist 302-42870-04 (Rev A) — Sep 2017...
Save all shipping cartons where equipment is to be declared as damaged or lost. The claims investigator must examine the cartons prior to completion of the inspection report. TriStar II 3020 Confirm Installation Instructions and Checklist 1 - 1 302-42870-04 (Rev A) — Sep 2017...
If the dewar has a protective mesh covering, do not remove it. This cover minimizes the risk of fly- ing particles should the dewar be knocked over, dropped, or broken. TriStar II 3020 Confirm Installation Instructions and Checklist 1 - 2...
2. Plug the other end of the Ethernet cable into the computer Ethernet port. ONNECT ULTIPLE NALYZERS WITH AN THERNET WITCH Do not use the uplink port when using an Ethernet switch. TriStar II 3020 Confirm Installation Instructions and Checklist 2 - 1 302-42870-04 (Rev A) — Sep 2017...
2. Check the power indicator on the front panel of the analyzer. The indicator will blink green to indicate that power is applied to the analyzer. Once the application is running, the indicator will stop blinking and remain illuminated. TriStar II 3020 Confirm Installation Instructions and Checklist 2 - 2 302-42870-04 (Rev A) — Sep 2017...
The person installing the application must have Windows Administrator rights.The administrator must create profiles for all users running Micromeritics TriStar II Confirm. The installation media contains the Micromeritics application and the Administrator Utility application. The following tasks must be completed prior to application installation: The Ethernet port for the analyzer must be set to 192.168.77.100...
Page 12
For example, if the application will only be installed on this computer, enter 01. If installation will also be done on a second computer, enter 02, etc. The TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 2...
Page 13
Click Next. 11. On the Ethernet Ports window, select one option from the following table: TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 3 302-42870-04 (Rev A) — Sep 2017...
Page 14
If installing multiple analyzers, at the prompt, insert the media containing the calibration files for the next analyzer. 13. On the Installation Complete window, read the notices, then click Finish. TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 4 302-42870-04 (Rev A) — Sep 2017...
Page 15
Upgrade applications do not contain calibration files. 15. If prompted, restart the computer after application installation. TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 5 302-42870-04 (Rev A) — Sep 2017...
Service Personnel have different directory and file access permissions. In addition to creating user groups, the Micromeritics application also sets the Windows directory and file access permissions for each of these user groups.The default file permissions are included in the Administrator Utility for TriStar II 3020 Confirm User Guide.
In addition to creating user groups, the Micromeritics application also sets the Windows directory and file access permissions for each of these user groups.The default file permissions are included in the Administrator Utility for TriStar II 3020 Confirm User’s Guide. These settings may be modified as needed.
The pass- word must contain a minimum of six alphanumeric characters. This password will be used to log in to the Administrator Utility and the Micromeritics application. This will not affect the Win- dows password.
Create Password Configuration Parameters REATE ASSWORD ONFIGURATION ARAMETERS The Password parameters entered here apply to all Micromeritics application users. On the Administrator Utility window, click the Configuration tab. Administrator Utility Fields Field Description Enter the number of days when a user password will expire. This num- Expire ber of days goes into effect any time a password is changed.
Enter password as a temporary password that can be used to access Password the application. This password is not saved in the password history file. TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 10 302-42870-04 (Rev A) — Sep 2017...
Page 21
Selected because the user’s assigned password is temporary and Change password on must be changed when the user logs in to the application. next logon 3. Click to save. TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 11 302-42870-04 (Rev A) — Sep 2017...
ERVICE CCOUNT A Service user account should be set up during installation to provide Micromeritics service personnel access to system functions if a service call is required. 1. On the Administrator Utility window, click the User tab, then click Add.
TART THE PPLICATION To start the application, click the application shortcut on the desktop or select the application from the Windows program list. TriStar II 3020 Confirm Installation Instructions and Checklist 3 - 13 302-42870-04 (Rev A) — Sep 2017...
Prior to the connection of the gas inlet lines to the analyzer, the manifolds are evacuated to prepare them for the installation of the gas lines. 1. Open the application and go to Start > All Programs > Micromeritics > TriStar II Confirm.
Page 26
7. To zero the transducers, go to Unit 1 > Calibrate > Zero Pressure. Select the four trans- ducers and click Start. Verify the pressure transducers have been zeroed. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 2 302-42870-04 (Rev A) — Sep 2017...
Use to specify how data should appear on the application windows and reports. This menu option is not available if using Restricted option presentation. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 3 302-42870-04 (Rev A) — Sep 2017...
4 Analyzer Preparation ONNECTIONS The TriStar II 3020 uses pressure relief valves to return the analyzer to a safe condition in the event of a power failure or loss of cryogen. This protects the user from dangerous pressures on the sample tube. However, extra caution should be employed if toxic or flammable gases are used.
Ensure the gas cylinder is closed before connecting to the analyzer. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 5 302-42870-04 (Rev A) — Sep 2017...
If pressure drops, repair the leak. 6. Connect the gas line to the regulator. a. Connect the gas line to the regulator connector. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 6 302-42870-04 (Rev A) — Sep 2017...
Page 31
Open the gas cylinder shutoff valve and flow gas for 10 to 30 seconds. c. Close the gas cylinder shutoff valve and allow regulator pressure to go to zero. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 7...
3. Use a 7/16 in (11 mm) wrench to tighten the nut an additional 1/4 turn. 4. Connect all the remaining gases in the same manner. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 8 302-42870-04 (Rev A) — Sep 2017...
3. Click the down arrow to the right of the Test field and select Clean and Verify Analysis Gas Line [number] Rev. [latest revision letter]. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 9 302-42870-04 (Rev A) — Sep 2017...
Page 34
If Failed is shown instead, one or more valves is not in the correct position. Set the valves as shown below: TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 10 302-42870-04 (Rev A) — Sep 2017...
Page 35
Clean and Verify the Gas Line 10. Click Close to close the test report. 11. Click Close on the test dialog to close the test. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 11 302-42870-04 (Rev A) — Sep 2017...
Micromeritics service representative. Board ID. Displays information from the electronic circuit boards in the analyzer. These parameters cannot be edited. TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 12 302-42870-04 (Rev A) — Sep 2017...
Page 37
Servo Valve [group box] Displays the last calibration date for the servo valve. Displays the software versions of the MIC BIOS, controller, and ana- Software Versions lysis program. [group box] TriStar II 3020 Confirm Installation Instructions and Checklist 4 - 13 302-42870-04 (Rev A) — Sep 2017...
Pressurize the manifold, sample ports and P port to 800 to 900 mmHg. b. Go to Unit [n] > Calibration >Pressure Scale. c. Select: Manifold Transducer TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 1 302-42870-04 (Rev A) — Sep 2017...
Page 40
If the system has a 10 mmHg transducer, continue with the filter frits removed, otherwise re-install the filter frits in the sample port 1 and 2. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 2 302-42870-04 (Rev A) — Sep 2017...
8. Record the serial number of the pressure calibration standard on the data sheet. 9. Remove the vacuum gauge and pressure calibration standards. 10. Re-install the filter frits in the sample port 1 &2. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 3 302-42870-04 (Rev A) — Sep 2017...
Page 42
(only if the reference temperature meter has been calibrated within 1 year.) 6. Go to Unit 1 > Calibration > Temperature. 7. Enter current temperature. Click OK. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 4 302-42870-04 (Rev A) — Sep 2017...
Report Options Click Advanced to view the contents of the selected files. 5. Click Save and complete the Save As options. 6. Click Close. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 5 302-42870-04 (Rev A) — Sep 2017...
Page 44
5. Click each parameter tab and make any necessary edits. Refer to the TriStar II 3020 Confirm Operator Manual for detailed instructions on these fields. 6. Click Save and complete the Save As options. 7. Click Close. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 6 302-42870-04 (Rev A) — Sep 2017...
Sample tubes and filler rods must be clean and dry before samples are added and weighed. The following table indicates which materials are supplied by Micromeritics and which are supplied by the user. The procedures following the materials table are recommended.
Page 46
12. Label the sample tube and stopper for identification. 13. Replace the rubber stopper, Check Seal, or TranSeal. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 8 302-42870-04 (Rev A) — Sep 2017...
8. Use the sample tube funnel (provided in the accessories kit) and pour the sample from the weighing container into the sample tube. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 9 302-42870-04 (Rev A) — Sep 2017...
2. Subtract the Mass for empty sample tube set (Before Degas) from the Sample tube set plus sample mass (After Degas) to obtain the sample’s mass; record this value as Sample mass (After Degas). TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 10 302-42870-04 (Rev A) — Sep 2017...
Ensure the tapered side of the cover is facing downward. 3. Insert the sample tube through one of the holes in the cover. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 11 302-42870-04 (Rev A) — Sep 2017...
Page 50
The equipment images in this document may differ slightly from your equipment, how- ever the instructions are the same unless otherwise noted. See also: Dewar Precautions on page 1 - 2 TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 12 302-42870-04 (Rev A) — Sep 2017...
4. If using isothermal jackets, slide the jackets down the sample tube until the jackets touch the sample tube bulbs. 5. Slide the dewar lid to ensure a proper seal on the top of the dewar. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 13 302-42870-04 (Rev A) — Sep 2017...
Page 52
3. Fill the dewar with LN and check the level using the dipstick. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 14 302-42870-04 (Rev A) — Sep 2017...
Volume adsorbed to be ±0.2000 std cm at any point. The isotherm 1/2 inch tubes must show a fairly smooth trace. No sudden peaks, except at saturation. TriStar II 3020 Confirm Installation Instructions and Checklist 5 - 15 302-42870-04 (Rev A) — Sep 2017...
Page 54
Cause: The measured free space is too high, indicating that the helium is not pure enough. Action: Use helium that is 99.999% pure, then try again. 8. Retain a copy of the sample information file to return to the Micromeritics Service Support Center.
Ensure that all equipment has been unpacked and verified using the packing list. Analyzer is located in a suitable environment Line voltage has been verified to be within tolerances and changed if required. TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 1 302-42870-04 (Rev A) — Sep 2017...
Page 56
(make / model / SN) Computer Operating System Printer (make / model / SN) Was the computer supplied by Micromeritics? Computer Installed By: Name: Date: Comments: TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 2 302-42870-04 (Rev A) — Sep 2017...
Page 57
Requirements for non-standard installations. Record issues con- sidered non-standard. Network prepared by Network Administrator, if applicable. File locations determined. TriStar II 3020 application installed. TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 3 302-42870-04 (Rev A) — Sep 2017...
Page 58
Install the Vacuum Pump Evaluation Initial / Procedure Code Date Install the components of the vacuum pump. Connect the vacuum pump to analyzer. TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 4 302-42870-04 (Rev A) — Sep 2017...
Page 59
Specify units of measure. Connect gas lines. Clean and verify gas lines. Specify Gases Evaluation Initial / Procedure Code Date Gases entered in application? TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 5 302-42870-04 (Rev A) — Sep 2017...
Page 60
Initial / Date Expected BET results: _______________±_______________ Y N N/A Actual BET results: ________________±________________ Port 3: Carbon Black Initial / Date Expected BET results: _______________±_______________ Y N N/A Actual BET results: ________________±________________ TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 6 302-42870-04 (Rev A) — Sep 2017...
Exceptions XCEPTIONS Explain the procedures that were categorized as Fail or No. Comment Initial / Date Procedure TriStar II 3020 Confirm Installation Instructions and Checklist 6 - 7 302-42870-04 (Rev A) — Sep 2017...
Service Support Center in Norcross, GA, USA. After completing the installation process, representatives of Micromeritics who install analyzers are required to send the following documents to Micromeritics for inclusion in the customer's analyzer history:...
UTOMATIC LANK ONDITIONS The TriStar II 3020 application contains an automated procedure of conducting the blank tube run. Use following steps to create blank tube run conditions and start the procedure. 1. Go to Unit 1 > Service Test > PCP Analysis.
Page 64
P0: 745.000 mmHg Temperature: 77.30 K Equilibration Equilibration interval: 6 sec Minimum equilibration delay at P/P0 > 0.995: 600 secs Do not adjust Backfill TriStar II 3020 Confirm Installation Instructions and Checklist A - 2 302-42870-04 (Rev A) — Sep 2017...
Plot Adsorption Branch Plot Description Branch Linear Plot Selected Plot Curve Plot Points Overlay Samples Autoscale X Axis Autoscale Y Axis Logarithmic Plot Selected Plot Curve TriStar II 3020 Confirm Installation Instructions and Checklist A - 3 302-42870-04 (Rev A) — Sep 2017...
Page 66
User Isothermal Jacket Use Filler Rod EGAS ONDITIONS ETTINGS Field Enter Degas Conditions Description Stage Soak Temperature (°C) Ramp Rate (°C/min) Soak Time (min) TriStar II 3020 Confirm Installation Instructions and Checklist A - 4 302-42870-04 (Rev A) — Sep 2017...
Lower dewar after free space: Yes Evacuation time: 0.10 hours Outgas test: No and T P0 type: Measure P0: 760.000 mmHg Temperature: 77.30 K TriStar II 3020 Confirm Installation Instructions and Checklist A - 5 302-42870-04 (Rev A) — Sep 2017...
Page 68
Autoscale X Axis Autoscale Y Axis Logarithmic Plot Selected BET Surface Area: Tabular Report Selected Plot Report Selected Overlay Samples Autoscale X Axis Autoscale Y Axis Langmuir Surface Area TriStar II 3020 Confirm Installation Instructions and Checklist A - 6 302-42870-04 (Rev A) — Sep 2017...
Page 69
Single Point BET Multi Point BET Equilibration 5 sec Equilibration interval 600secs Minimum equilibration delay at P/P0 ≥ 0.995 Do not adjust Backfill TriStar II 3020 Confirm Installation Instructions and Checklist A - 7 302-42870-04 (Rev A) — Sep 2017...
Worksheets in this section may be copied as needed. See also: Sample Data Worksheet on the next page User Information Worksheet on page B - 3 TriStar II 3020 Confirm Installation Instructions and Checklist B - 1 302-42870-04 (Rev A) — Sep 2017...
C. Sample mass (B - A): Degas Information Degas apparatus: Temperature (ºC): Time (hours): Actual time started: Actual time finished: Degas Notes: TriStar II 3020 Confirm Installation Instructions and Checklist B - 2 302-42870-04 (Rev A) — Sep 2017...
* User Name: Windows User ID ** Service: Users should be added to the MicService Windows User Group and assigned a Developer User Profile in the Administrator Utility. TriStar II 3020 Confirm Installation Instructions and Checklist B - 3 302-42870-04 (Rev A) — Sep 2017...