ASAP 2020 Confirm
Calculate
Po and T
Displays as Analysis Bath
Temperature only when
Absolute dosing is
selected (Analysis dialog)
4-24
Select this option to have the free-space calculated using the
mass, density, and sample tube parameters entered in the
Sample Information and Sample Tube files. Calculated free
space measurement takes less time than measured free
space, so if you know the mass and density of the sample
and the sample tube parameters, you may wish to use this
choice of free space methods. If you do not know the
parameters, you may wish to use measured free space.
Displays the Po and Temperature Options dialog, allowing
you to choose saturation pressure and analysis bath
temperature. A field is also provided for entering the ambient
temperature.
This dialog provides six options for obtaining the saturation
pressure (Po) and analysis bath temperature. Clicking on each
option prompts you for the entry of related parameters in the
lower portion of the dialog.
Option
Prompts for...
1
a measurement interval (how often during
analysis the Po is to be measured)
2
a measurement interval and the analysis
bath temperature
3
a Po; the analysis bath temperature is
calculated at the time of analysis
New
202-42811-01 - Mar 2011