Displays the Entered t-Curve dialog allowing you to define a
t-curve by entering the relative pressure and thickness values.
This table can also be created in another application if desired
and imported into this dialog using
table in another application, the file must be saved as ASCII
text with a. THK extension. Use a two-column format with
the relative pressures in the first column and the thickness
values in the second column. Columns must be separated by a
space (or a tab).
Insert
Inserts a row above the selected row. A row cannot be
inserted if the value in the selected row is at its lowest value;
values must be strictly increasing. For example, in the table
shown above, the default value is 0.000000001. Therefore
you must use
Ctrl + Down Arrow
Delete
Deletes the selected row.
Clear
Clears the table of all but one entry, one entry is required.
Open
Enables you to import the values from an existing t-curve.
Save As
Enables you to save the current table of values under a file
name.