No data points are within the range of pressures in the reference isotherm.
Cause:
Action:
4068-
No points were selected for the f-Ratio report.
Cause:
Action:
4069-
Dosing method choice is invalid. The Analysis Conditions choice of Absolute
pressure dosing requires that the Adsorptive Properties (Dosing Method) is set to
Normal.
Cause:
Action:
4070-
Unable to load deconvolution model (name).
Cause:
Action:
B-40
Thre are no collected data points within the range of pressures in the
reference isotherm.
Select data points that are in the range of the reference isotherm, or
select a more appropriate reference isotherm.
The collected data column for the f-Ratio report does not have any
points selected.
Edit the collected data dialog and select points for the f-Ratio report.
There is an incompatibility between the analysis conditions choice of
Absolute pressure dosing and the adsorptive properties dosing
method.
Edit one of the choices.
For some reason, the list of available models was corrupted, therefore,
the model selected could not be loaded for the deconvolution.
Exit the program and reinstall the software, then try again.