Micromeritics ASAP 2020 Confirm Operator's Manual page 97

Developer/analyst
Table of Contents

Advertisement

New
Absolute pressure dosing
Use calculation
assignments
Preparation
Fast evacuation
Unrestricted evac from
Vacuum setpoint
202-42811-01 - Mar 2011
Select this option to specify pressure targets in mmHg, mbar,
or kPa instead of relative pressure. This option is typically
selected when using adsorptives at analysis conditions above
the critical point of the gas; for example, H
carbon at liquid nitrogen temperature.
Select this option to assign the points to be collected for each
type of report.
If this option is not selected, each report (with the exception
of Langmuir and BET) uses a range of pressures as selected
in the report options. The Langmuir and BET reports
interpolate to entered relative pressures on the report options.
The outlier points can be selected so that they will not be
reported.
Displays the Analysis Preparation dialog box.
Select this option for samples (such as pellets) that do not
fluidize or shed particles during evacuation.
Enabled when Fast evacuation is not desired. This field
allows you to enter the pressure at which unrestricted sample
evacuation is to begin.
Allows you to specify the vacuum level to be achieved before
timed evacuation begins.