Preventive Maintenance
4. Remove the traps by opening both vacuum clamps and separating the trap from the
vacuum pump and hose connections.
5. Remove one end fitting from each trap body.
6. Remove the used alumina from the trap body and dispose of it in an appropriate manner.
202-42811-01 - Mar 2011
Centering Ring
Clamp
Clamp
Centering Ring
Hose
Hose Fitting
Oil Vapor Trap
(optional)
Intake
End Fitting
O-ring
Body
O-ring
End Fitting