ASAP 2020 Confirm
Temkin Isotherm
The Temkin isotherm is used to model adsorption data where the heat of adsorption drops
linearly with increasing coverage.
Specify monolayer
capacity
Specify differential
heat of adsorption
at zero surface coverage
Absolute pressure range
Tabular report
Transform plot
4-48
In this field, enter the monolayer capacity of the sample.
Enter the differential heat of adsorption at zero surface
coverage. This allows inclusion of all of Temkin constants.
Allows you to enter a pressure range when calculation
assignments are not requested.
Select this option to have a tabular report of the pressure
points generated
Plots a linear form of the Temkin isotherm.
New
202-42811-01 - Mar 2011
Need help?
Do you have a question about the ASAP 2020 Confirm and is the answer not in the manual?