Installing the ASAP 2420 1-mmHg Transducer Option (Units with Serial Number 199 and below) P/N 242-42805-00 This document contains instructions for installing the 1-mmHg Transducer option only or for installing both the 1-mmHg transducer and micropore options on an ASAP 2420 analyzer.
Page 2
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option 5. Lift the panel up and off the analyzer. 6. The 10-mmHg transducer must be removed from the top of the old transducer block and installed on the front of the new block, and the 1-mmHg transducer must be installed on top of the new block.
Page 3
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only 7. To remove the 10-mmHg transducer, first disconnect the 10-mmHg transducer ribbon cable from the J8 connector on the transducer board (P/N 202-17706-001). 8. Use a wrench to loosen the nut that connects the 10-mmHg transducer to the transducer block.
Page 4
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option 10. Use a wrench to loosen the nuts that connect the transducer block to the manifold tee connectors and to the gas inlet line, then remove the block. Remove connector nuts 11.
Page 5
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only 12. Assemble the new transducer block, spacers, ferrules and connector nuts. Screw Front and rear ferrules (under nut) Lock washer Spacer Connector nut Rear ferrule Front ferrule New Block Use new spacers, ferrules, screws and lock washers from the upgrade kit.
Page 6
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option 15. Place the connector nut, and front and rear ferrules on the 1-mmHg transducer. Front ferrule Rear ferrule 16. Install the transducer on top of the transducer block; tighten securely with a wrench.
Page 7
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only 18. Install the 10-mmHg transducer, with new front and rear ferrules, in the front of the transducer block. Use the connector nut you removed earlier; tighten the nut securely with a wrench.
Page 8
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option 20. Attach the ribbon cable from the 10-mmHg transducer to the J8 connector on the transducer board. 10-mmHg transducer 21. Replace the top panel. 22. Replace and tighten the top panel retaining screws.
Page 9
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only 26. Select Unit [n] > Show Instrument Schematic > Enable Manual Control. 1-mmHg transducer 10-mmHg transducer 27. Open valves 7 and 8. Make sure the 1-mmHg and 10-mmHg transducer readings are displayed on the schematic.
Page 10
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options This set of instructions describes how to install the 1-mmHg Transducer option and the Micropore option on ASAP 2420 Krypton systems.
Page 11
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options 6. The 10-mmHg transducer must be removed from the top of the old transducer block and installed on the front of the new block, and the 1-mmHg transducer must be installed on top of the new block.
Page 12
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option 8. Use a wrench to loosen the nut that connects the 10-mmHg transducer to the transducer block. Retain the nut, you will use it to reinstall the transducer...
Page 13
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options 11. Remove the block mounting screws, connector nuts, and the spacer and ferrules from each tee connector. Screw Front and rear ferrules (under nut) Lock washer Spacer Connector nut...
Page 14
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option 14. Use a wrench to tighten the nuts that connect the transducer block to the manifold tee connectors and to the gas inlet line. Tighten connector nuts 15. Place the connector nut, and front and rear ferrules on the 1-mmHg transducer.
Page 15
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options 16. Install the transducer on top of the transducer block; tighten securely with a wrench. 17. Connect the ribbon cable to the 1-mmHg transducer and to J7 on the transducer board (202-17706-001).
Page 16
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option Make sure the zero adjustment on the 10-mmHg is accessible. You may have to cut away some of the foam if it interferes with the transducer. Zero adjustment 19.
Page 17
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options 20. Plug the Micropore key into the J10 connector on the transducer board. Use a screwdriver to tighten the retaining screws. Micropore key (J10) 21. Replace the top panel.
Page 18
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option 26. Remove the filter and O-ring. O-ring Filter 27. The filter and O-ring can be used with the new port fittings unless either is damaged. A damaged filter or O-ring should be replaced with a new one.
Page 19
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options 33. Select Unit [n] > Show Instrument Schematic > Enable Manual Control. 1-mmHg transducer 10-mmHg transducer 34. Open valves 7 and 8. Make sure the 1-mmHg and 10-mmHg transducer readings are displayed on the schematic.
Page 20
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option Feb 2010...