7. Using a funnel, pour sample from the container into the sample tube.
If some sample clings to the inside of the sample tube above the last 3 in. (7.5-cm) of the
tube, use a pipe cleaner or lint-free wipe to remove it.
8. Replace the filler rod and insert the rubber stopper or seal frit.
9. Weigh the sample tube set containing the sample and record the weight on the Sample
Data Worksheet as Mass of sample tube plus sample (Before Degas).
10. Subtract the Mass of empty sample tube from the Mass of sample tube plus sample; record
this value as the Mass of sample.
3-20
Preparing Samples
202-42811-01 - Mar 2011