ASAP 2020 Confirm
4015-
Error creating export file for sample <sample file name>.
Cause:
Action:
4016-
Sample (file name) has no data for export.
Cause:
Action:
4017-
Damage to the instrument will result if the sample has not been manually evac-
uated. Have you evacuated the sample?
Cause:
Action:
B-28
A file error occurred during creation of an export output file.
The output file name may be invalid. Ensure that the target directory
exists. Ensure that the target diskette is not full or write protected. The
target disk drive may be damaged or inoperative. Verify that other files
may be created on the same drive. Contact your service representative
if you continue to receive this error message.
The file selected for export has a status of No Analysis. No export file
will be created.
Select a file which contains analysis data.
You did not select Backfill sample at start of analysis on the Sample
Backfill Options dialog box. The sample tube is normally at atmo-
spheric pressure when an analysis is started, and it must be backfilled
before the analysis begins to prevent sample material from being
drawn into the manifold.
If you have manually evacuated the sample tube, select
have not, select
and then either perform a manual evacuation or go
No
to the Sample Backfill Options dialog box and select Backfill sample
at start of analysis.
Appendix B
. If you
Yes
202-42811-01 - Mar 2011