ASAP 2020 Confirm
Horvath-Kawazoe Report Options
This report is available only if you have the Micropore option installed.
Pore Geometry
Apply Cheng/Yang
correction
Pressure Range
Interaction Parameter
4-70
Select the option in this group box which best represents the
physical geometry of the micropores in the sample material.
Refer to Horvath-Kawazoe, page
information on calculations used for these parameters.
Allows you to apply the Cheng/Yang correction to the pore
size analysis. This correction substitutes the Langmuir
equation of state for Henry's Law in the Horvath-Kawazoe
derivation (refer to CALCULATIONS, page C-1).
Displays the Report Relative Pressure Range dialog so that
you may specify minimum and maximum relative pressures
to use with this report.
• If Use calculation assignments is not selected on the Col-
lected/Entered dialog, all of the nonoutlier points of the
collected data within the specified range are used for calcu-
lating the data for this report.
• If Use calculation assignments is selected, collected data
points which are assigned to this report type are used.
Select one of the options in this group box to determine
which interaction parameter is used during report generation.
New
C-22
for additional
202-42811-01 - Mar 2011
Need help?