Preventive Maintenance
Performing a Leak Test
This procedure may be performed at the request of your service representative to determine if
there is a leak in your system This test generates a report that your service representative will
request that you send to him for observation.
The following types of information are displayed during the test:
•
Prompts on preparing the instrument for the test
•
Approximate time period of the test
•
Prompts in which an operator response is required
A manual procedure for checking leaks is also available. Refer to
LEAKS, page
1.
Select Unit [n] > Diagnostics; the Service Test dialog is displayed.
2. Click the down arrow to the right of the Test field and select System Leak Test Rev.
[latest revision letter].
3. Ensure that Report after test is selected and that Screen is chosen as the destination.
202-42811-01 - Mar 2011