Appendix B
6531-
The gas configuration file for (unit number) is invalid.
Cause:
Action:
6532-
The instrument is currently running the Physi software. Do you want to reset
it?
Cause:
Action:
6533-
The instrument is currently running the Chemi software. Do you want to reset
it?
Cause:
Action:
6534-
Instrument (unit number) is not calibrated.
Cause:
Action:
6535-
Problem encountered evacuating.
Cause:
Action:
202-42811-01 - Mar 2011
The contents of the gas configuration file are not valid.
Confirm there is sufficient free space on the media receiving the file
and that the media is not corrupted (run ScanDisk). If the problem per-
sists, contact a Micromeritics service representative.
The analyzer is under the control of the 2020 Physisorption software.
Selecting
will reset the analyzer and download the 2020 Chemi-
Yes
sorption control software.
The analyzer is under the control of the 2020 Chemisorption software.
Selecting
will reset the analyzer and download the 2020 physi-
Yes
sorption control software.
Calibration information for various analyzer components are missing.
Run the application setup program and reinstall calibration information
for the specified unit.
Evacuation during the manifold dosing operation did not come within
the allowed range of the target.
Check that the outlet stage of the gas regulator is within specification.
If the problem occurs frequently, contact a Micromeritics service rep-
resentative.
ASAP 2020 Confirm
B-55