Appendix B
4060-
Sample (file name) contains no data points.
Action:
Cause:
4061-
The t-curve must contain at least 2 points.
Cause:
Action:
4062-
Error during report generation.
Cause:
Action:
4063-
The data requested on this report are not available.
Cause:
Action:
4064-
This report cannot be produced without the Micropore option.
Cause:
Action:
202-42811-01 - Mar 2011
An attempt was made to save a sample without collected data as a t-
curve or alpha-S curve.
Repeat the save as t-curve or save as alpha-S operation after opening a
sample that has collected data.
At least two points are required in a thickness curve definition.
Edit the thickness curve.
An internal processing error has occurred.
Contact your Micromeritics service representative.
There is no information in the sample log to report.
You selected a sample file for which no instrument operations have
been used. Select a sample file with a status of Prepared, Preparing,
Analyzing, or Complete to obtain a valid sample log report.
You requested a report that requires the micropore option.
Deselect the report. Order the Micropore option to install on teh ana-
lyzer.
ASAP 2020 Confirm
B-39