Micromeritics ASAP 2020 Confirm Operator's Manual page 115

Developer/analyst
Table of Contents

Advertisement

New
Pass/Fail
Upper/Lower
202-42811-01 - Mar 2011
Displays the Pass/Fail Options dialog so that you can choose
a parameter on which to specify pass/fail criteria.
Select these options to specify upper and lower limits for the
selected parameter; a valid range for the selected field is
displayed in the information bar.
You can leave the range open by deselecting one of the limits.
For example; if you wish to leave the upper limit open,
deselect the Upper check box and specify a value only for
the Lower limit.
Each of these fields has a message line in which you can
enter advice to the operator if a failure occurs. You may enter
up to 120 characters in each field.