ASAP 2460 Operator's Manual
4000 Series
4002-
Thermal Transpiration correction had no effect.
Cause:
Action:
4003-
Error Converting Pressures.
4004-
Error Computing Quantity Adsorbed.
Cause:
Action:
4005-
Pressures were not smoothed. Not enough pressures below 0.10 P/Po.
Cause:
Action:
4011-
Analysis gas in sample file does not match analysis gas in unit.
Cause:
Action:
246-42800-01 - Dec 2013
The Thermal transpiration correction option was selected on the Report
Options dialog box. However, the correction did not change any pres-
sure by more than one percent.
De-select this option to disable this message. This correction is only
meaningful for very low pressures.
An internal processing and/or hardware error occurred during report
generation.
Contact your service representative if you continue to receive this error
message.
The Smooth pressures below 0.10 P/Po option was selected on the
Report Options dialog. There must be at least 10 pressures within this
range for smoothing to occur.
Deselect this option to disable this message.
You attempted to start an analysis using a sample information file in
which the analysis gas specified does not match the analysis gas
entered in the unit configuration.
If necessary, attach the appropriate gas bottle, then enter the gas in the
Unit Configuration dialog box.
Appendix B
B-17