ASAP 2020 Confirm
Performing an Analysis
After the sample has been degassed and transferred to an analysis port, you may begin an
analysis.
You cannot perform an analysis while calibration is in progress.
Before beginning an analysis, make sure the tank pressure for the gas regulator is at
least 200 psig. Pressures less than 200 psig may cause the sample to be inadequately
saturated, resulting in inaccurate data or termination of analysis.
Ensure that the analysis gas and the Psat gas specified in the sample file match the system
configuration gases; if it doesn't, correct the sample information file or correct the unit
configuration gas.
Perform the following steps to begin an analysis:
1. Select Sample Analysis from the Unit menu; the Analysis dialog is displayed with the
Start Analysis dialog positioned on top.
2. Choose a file for your analysis and click
parameters of the selected file is displayed.
3. Verify parameters and make any changes you feel necessary.
3-28
; the Analysis dialog containing the
OK
Performing an Analysis
202-42811-01 - Mar 2011
Need help?
Do you have a question about the ASAP 2020 Confirm and is the answer not in the manual?