6500- Failed to evacuate manifold to VAC SET in (number of) seconds. Calibration
canceled.
Cause A:
Action A:
Cause B:
Action B:
Cause C:
Action C:
6501-
The 1000 mmHg transducer offset exceeds recommended limits: (number)
Cause:
Action:
6502-
The 10 mmHg transducer offset exceeds recommended limits: (number)
Cause:
Action:
202-42811-01 - Mar 2011
The vacuum set point is set too low.
Ensure that the vacuum set point is at 5.0 mmHg. If the set point is
already at 5.0 mmHg or above, the vacuum gauge may need servicing.
Leak in manifold.
Locate the leak and repair it. Refer to Chapter 8, Troubleshooting and
Maintenance. Restart calibration.
Valve failure.
Identify the leaking valve. Contact your service representative.
The Pressure Gauge Calibration operation showed the transducer offset
exceeds the recommended limit.
Repeat the Pressure Gauge Calibration operation, if this message
occurs again contact a Micromeritics service representative.
The Pressure Gauge Calibration operation showed the transducer offset
exceeds the recommended limit.
Repeat the Pressure Gauge Calibration operation, if this message
occurs again contact a Micromeritics service representative.