After setting test parameters as described in
page
9, you can start your EMF measurements.
To set measure setup
1
Tap the Setup (
2
Tap Test Configuration.
a
b
c
3
Tap to switch Axis to Manual or Auto.
If you set it to Manual, Axis setting is available and if you set it to Auto, Axis
setting is unavailable.
4
When selected as Manual, tap Axis and select from: X, Y, or Z.
5
Tap Move to RtSA to view the real-time spectrum measurement mode.
April 2021
) icon in the side bar.
Tap to switch Auto Range to On or Off.
When you set Auto Range On, the instrument scans to-be-measured band-
width for EMF in advance and sets appropriate Attenuation and Preamp
based on the different received signal level for the choice of axis.
Tap Dwell Time to specify the amount of measurement time that you want
the instrument to stay for on each axis and input the value from 1 to 60
(second) using the on-screen keyboard.
Tap Measurement Time to specify the amount of measurement time that you
want the instrument to stay for on all three axes in order to display current
and accumulated EMF power and set this up to 60 minutes using the on-
screen keyboard.
NOTE
If you set the measurement time, the count which means the number of EMF
measurements on the right panel of the measurement screen is changed cor-
responding to the measurement time you have set.
Manual: When using isotropic EMF antenna, you can manually set x, y, and z
axis. You can set Manual as a diagnosis mode whether you can receive
correct signal based on each axis: X, Y, or Z.
Auto: Each axis of antenna is automatically changed and measured.
SPA06MA-O User's Guide
"Configuring spectrum measurements" on
22145360, R005
Chapter 11 Using EMF Analyzer
Conducting spectrum measurement