SSB SCS (kHz)
30
120
240
To set limit
1
Tap Menu > Limit in the side bar.
2
Tap Standard Limit Line and select from the following options:
Standard Line is displayed on the screen based on the pre-defined standard. If
you select Occupational or General Public, the Standard Line is automatically
displayed with automatically calculated value defined in the ICNIRP guideline.
Make sure you change the Scale Unit to V/m in this case to follow the guideline.
a
b
c
To set sweep
1
Tap Menu > Sweep in the side bar.
April 2021
Sync Raster
Resolution (kHz)
30
120
240
Tap Occupational and Apply if you are under impact on EMF power for an
occupational purpose.
Tap General Public and Apply if you are under impact on EMF power as
general public.
Tap Custom Limit and then tap the Configuration icon if you want to
customize setting and measure EMF power based on your need.
-
A Standard window appears.
-
Tap Lower Frequency and input a desired value using the on-screen
keyboard.
-
Tap Upper Frequency and input a desired value using the on-screen
keyboard.
-
Tap Value for Frequency Range and input a desired value using the on-
screen keyboard.
Tap Formula and select from the following choices: f^n, /f^n, NA
-
-
Tap Exponent Value and input a desired value using the on-screen
keyboard.
-
Tap the Save button if you want to save the setting.
-
Optional. Tap the Add button to add Custom Limit.
-
Optional. Tap the Delete button to delete the defined Custom Limit.
NOTE
If you set Formula to f^n or /f^n, you can define the limit line as frequency of
function. If you set it to NA, you can define the limit line as a constant.
SPA06MA-O User's Guide
Sync Raster
Offset Range
0 to 5
0 to 5
0 to 2
22145360, R005
Chapter 11 Using EMF Analyzer
Conducting signal measurement
SSB Frequency
= Center
Frequency
0
0
0
Page 389