Setting measure setup
After configuring the spectrum measurement in
on page
To set measure setup
1
Tap the Setup (
2
Tap Bandwidth and select the value from the following choices: 1.4 MHz, 3 MHz,
5 MHz, 10 MHz, 15 MHz, or 20 MHz.
You can also tap the rectangle with value,(
the chart screen and input the value using the on-screen keyboard.
3
Tap the Center Frequency (
screen to set the center frequency using the on-screen keyboard.
4
Tap Menu > BW/AVG > Average to set the number of measurements to be aver-
aged using the on-screen keyboard.The input value range is from 1 to 100.
To set limit
1
Tap Menu > Limit.
2
Tap to switch the Test Limits to On or Off to enable or disable Pass/Fail indica-
tion.
The result table color for Pass is green, and the result table color for Fail is red.
3
Tap High Limit to set the upper threshold using the on-screen keyboard.
4
Tap Low Limit to set the lower threshold using the on-screen keyboard.
5
Optional: Tap the Save hot key on the side bar and select the save option from the
choices available for your measurement mode.
Figure 39
April 2021
9, you can set the measure setup to continue your measurement.
) icon in the side bar.
shows an example of LTE/LTE-A FDD channel power measurement.
SPA06MA-O User's Guide
"Configuring spectrum measurements"
22145360, R005
Chapter 5 Using LTE/LTE-A FDD Analyzer
Conducting RF measurement
)icon under the chart
) under
Page 87