Viavi SPA06MA-O R005 User Manual page 192

Table of Contents

Advertisement

Chapter 6 Using LTE/LTE-A TDD Analyzer
Conducting OTA measurement
To set measure setup
1
Tap the Setup (
2
Tap Bandwidth and select the value from the following choices: 1.4 MHz, 3 MHz,
5 MHz, 10 MHz, 15 MHz or 20 MHz.
3
Tap Subframe Number to set the number of subframe using the on-screen
keyboard.
4
Tap Position to set the number of positions using the on-screen keyboard when
Time Cursor is set to On.
5
Optional. Tap Physical Cell ID and select the type between Manual and Auto.
6
Optional. To set the threshold for PDSCH, tap PDSCH Threshold and enter a
value by using the on-screen keyboard.
7
To set the time cursor at a specific position, tap Time Cursor and select On.
8
TTo start a new measurement, tap Reset.
9
Optional. To select the number of antenna ports, tap the MIMO and select 2x2 or
4x4.
10 Optional. To select the Cyclic Prefix, tap Cyclic Prefix and select Normal or
Extended.
Figure 90
Page 176
) icon in the side bar.
Auto lets the instrument detect the Physical Cell ID for the LTE signal auto-
matically.
Manual sets a specific Physical Cell ID for the LTE signal manually in order to
speed up with the synchronization with a BTS.
NOTE
For MBMS testing, this subframe number must be either 0 nor 5 as these
subframes 0 and 5 are not available for MBMS and it must be set to the MBMS
transmitted channel's subframe number.
shows an example of LTE/LTE-A TDD OTA datagram measurement.
SPA06MA-O User's Guide
22145360, R005
April 2021

Advertisement

Table of Contents
loading

This manual is also suitable for:

Spa06ma-oSpa06ma

Table of Contents