Viavi SPA06MA-O R005 User Manual page 51

Table of Contents

Advertisement

Figure 17 Gate Sweep Spectrum Setup screen
To set Gate Sweep Spectrum
1
Tap Gate Delay on the setup table and input a value using the on-screen
keyboard.
When the gate delay changes, the gate length line also changes accordingly with
the same interval.
2
Tap Gate Length on the setup table and input a value using the on-screen
keyboard. The gate length cannot be moved on the left side of the gate delay.
3
Tap Zero Span Time on the setup table and input a value using the on-screen
keyboard.
4
Tap StdSignal to select a Standard Signal from the pop-up list.
5
Tap Period and switch it to Standard or Manual and input a value using the on-
screen keyboard.
6
Tap the X button on the setup table. Gate Sweep becomes On.
The setup table disappears and the edit mode icon appears on the right edge of
the chart screen, as shown in
Figure 18 Gate Sweep Spectrum with spectrum analyzer
7
When the judged Triggering is not valid, for example a Delay set larger than
Trigger Period, the "Trigger Not Detected" message appears on the screen.
April 2021
Figure
18.
SPA06MA-O User's Guide
22145360, R005
Chapter 2 Using Spectrum Analyzer
Page 35