Figure 79 LTE/LTE-A TDD control channel measurement

Subframe

The Subframe measures the modulation accuracy of all the data and control channels
at the specified subframe (1 ms).
Setting measure setup
After configuring the spectrum measurement in
on page
measurement settings can be saved and recalled as a file by selecting File Type as
Setup and load the file onto the instrument using the Save/Load icons on the side bar.
To set measure setup
1
Tap the Setup (
2
Tap Bandwidth and select the value from the following choices: 1.4 MHz, 3 MHz,
5 MHz, 10 MHz, 15 MHz, or 20 MHz.
3
Tap Subframe Number to set the number of subframe using the on-screen
keyboard.
4
Optional. Tap Physical Cell ID and select the type Manual or Auto.
April 2021
9, you can set the measure setup to continue your measurement. The
) icon in the side bar.
NOTE
For MBMS testing, this subframe number must be either 0 nor 5 as these
subframes 0 and 5 are not available for MBMS and it must be set to the MBMS
transmitted channel's subframe number.
Auto lets the instrument detect the Physical Cell ID for the LTE signal auto-
matically.
Manual sets a specific Physical Cell ID for the LTE signal manually in order to
speed up with the synchronization with a BTS.
SPA06MA-O User's Guide
"Configuring spectrum measurements"
22145360, R005
Chapter 6 Using LTE/LTE-A TDD Analyzer
Conducting modulation measurement
Page 159