Chapter 2 Using Spectrum Analyzer
Conducting RF measurements
5
Tap Configuration to configure up to five offsets. The configuration table
appears, as shown in
You can also tap the Configuration icon under the chart screen.
Figure 12 Multi-ACP offset configuration table
a
b
c
6
Tap Span to set the frequency range over whit the instrument will sweep using the
on-screen keyboard.
7
Tap Menu > Sweep > Gated Sweep to set the Gated Sweep On or Off.
To set limit
1
Tap Menu > Limit.
2
Tap to switch the Test Limits to On or Off to enable or disable Pass/Fail indica-
tion.
The result table color is green for Pass and red for Fail.
3
Optional: Tap the Save icon on the side bar and select the save option from the
choices available for your measurement mode.
Figure 13 Multi-adjacent channel power measurement with spectrum analyzer.
Page 30
Figure
Select Offset from 1 to 5 and switch each to On to display or Off to hide the
active offset.
Select Offset Frequency to set the offset frequency using the on-screen
keyboard.
Select Upper to set the threshold for the power difference between the main
channel bandwidth and the selected offset bandwidth on the right using the
on-screen keyboard.
NOTE
Make sure that RM is selected when you perform the multi-adjacent channel
power measurement.
SPA06MA-O User's Guide
12.
22145360, R005
April 2021