Chapter 7 Using 5G NR Signal Analyzer
Conducting RF measurements
Setting measure setup
After configuring the spectrum measurement in
on page
measurement settings can be saved and recalled as a file by selecting File Type as
Setup and load the file onto the instrument using the Save/Load icons on the side bar.
To set measure setup
1
Tap the Setup (
2
Tap BS Type and select the option from 1-C/1-H, 1-O, or 2-O.
Set 1-C (Conducted)/1-H (Hybrid) for multi-band operation with mapping of trans-
ceivers to one or more antenna connectors (1-C) or TAB (transceiver array
boundary) connectors for 1-H. Set 1-O when BS operates at FR1 and 2-0 when
BS operates at FR2.
3
Tap Measure Type and select the option from Transmitted or Receiver.
4
Tap to switch Setting to User if you want user defined setting and tap Configura-
tion or the Configuration icon.
a
b
c
d
5
Tap to switch Setting to 3GPP if you want 3GPP standard defined setting and tap
Configuration or Configuration.
a
b
6
Tap Measurement Type between Full and Examine.
The Examine mode displays only the selected range while the Full mode lets the
instrument automatically change the selected range from one another.
7
Tap Average on the box of upper screen and enter the value between 1 and 100
to set the number of measurements to be averaged.
You can also access this menu through Menu > Average on the side bar.
Page 210
9, you can set the measure setup to continue your measurement. The
) icon in the side bar.
Tap Range under the chart screen and switch to On to display or Off to hide
the selected range in the result table.
You can select the range number between 1 and 20 to add as a new or
change the existing settings.
Tap Start Frequency/Stop Frequency and enter the value for the selected
range using the on-screen keyboard.
Tap Start Limit/Stop Limit and enter the lower limit/upper limit for Pass/Fail
indication.
Tap Attenuation/RBW/VBW and specify or select the value.
Tap Range under the chart screen and switch to On to display or Off to hide
the selected range in the result table.
You can select the range number between 1 and 20 to add as a new or
change the existing settings.
Tap Attenuation and specify or select the value.
SPA06MA-O User's Guide
"Configuring spectrum measurements"
22145360, R005
April 2021