Setting measure setup
After configuring the spectrum measurement in
on page
measurement settings can be saved and recalled as a file by selecting File Type as
Setup and load the file onto the instrument using the Save/Load icons on the side bar.
To set measure setup
1
Tap the Setup (
2
Tap Bandwidth from the following choices: 5 MHz, 10 MHz, 15 MHz, or 20 MHz.
3
Tap to switch Signal Tech Type to TDD or FDD.
4
Tap Menu > BW/AVG > Average to set the number of measurements to be aver-
aged using the on-screen keyboard.
The input value range is from 1 to 100.
To set the limit
1
Tap Menu > Limit.
2
Tap to switch the Test Limits to On or Off to enable or disable Pass/Fail indica-
tion. The result table color for Pass is green, and the result table color for Fail is
red.
3
Tap High Limit to set the upper threshold using the on-screen keyboard.
4
Tap Low Limit to set the lower threshold using the on-screen keyboard.
5
Optional. Tap the Save icon on the side bar and select the save option from the
choices available for your measurement mode.
Figure 137
Figure 137 5G DSS channel power measurement
April 2021
9, you can set the measure setup to continue your measurement. The
) icon in the side bar.
shows an example of channel power measurement.
SPA06MA-O User's Guide
"Configuring spectrum measurements"
22145360, R005
Chapter 9 Using 5G DSS Signal Analyzer
Conducting RF measurements
Page 293