Chapter 7 Using 5G NR Signal Analyzer
Conducting RF measurements
Setting measure setup
After configuring the spectrum measurement in
on page
measurement settings can be saved and recalled as a file by selecting File Type as
Setup and load the file onto the instrument using the Save/Load icons on the side bar.
To set measure setup
1
Tap the Setup (
2
Tap Bandwidth/SSB SCS and select the Setting in the pop-up window based on
your need. SSB is abbreviation for Synchronization Signal Block or SS Block and
it refers to Synchronization PBCH block since the synchronization signal and
PBCH channel are packed as a single block that always moves together. SSB
should be detected first in 5G NR frame. See
criteria based on the operating frequency.
3
Tap BS Type and select the option from 1-C/1-H, 1-O, or 2-O.
Set 1-C (Conducted)/1-H (Hybrid) for multi-band operation with mapping of trans-
ceivers to one or more antenna connectors (1-C) or TAB (transceiver array
boundary) connectors for 1-H. Set 1-O when BS operates at FR1 and 2-0 when
BS operates at FR2.
Table 39 Setup per operating frequency
Operating
frequency
< 3 GHz
(FR1)
< 3 GHz
(FR1)
< 3 GHz
(FR1)
3 to 6 GHz
(FR1)
3 to 6 GHz
(FR1)
3 to 6 GHz
(FR1)
> 6 GHz
(FR2)
Page 208
9, you can set the measure setup to continue your measurement. The
) icon in the side bar.
L
4
4
4
8
8
8
64
SPA06MA-O User's Guide
"Configuring spectrum measurements"
Table 39
Case
A
{2,8} +14*n
B
{4,8,16,20}
+28*n
C
{2,8} +14*n
A
{2,8} +14*n
B
{4,8,16,20}
+28*n
C
{2,8} +14*n
D
{4,8,16,20}
+28*n
22145360, R005
below for the setting
Slots in a
SCS
burst period
15 kHz
5 (1 ms)
30 kHz
10 (0.5 ms)
30 kHz
10 (0.5 ms)
15 kHz
5 (1 ms)
30 kHz
10 (0.5 ms)
30 kHz
10 (0.5 ms)
120 kHz
40 (0.125 ms)
April 2021