Chapter 7 Using 5G NR Signal Analyzer
Conducting spectrum measurements
Table 26 Sync SCS offset range per SSB SCS and sync SCS resolution
SSB SCS (kHz)
15
30
120
240
To set trigger
1
Tap the Menu > Trigger.
2
Tap Trigger and select the options from Internal, External, or GPS.
3
Tap Freq. Ref.(Frequency Reference) to set the clock source.
Figure 98
Figure 98 5G NR trigger spectrum measurement
Page 192
Sync Raster
Resolution (kHz)
15
30
120
240
Internal: Uses a 10 MHz internal clock
External: Uses external 10 MHz, 13 MHz, or 15 MHz clock automatically set
by the instrument
GPS: Uses a built-in GPS as a frequency and timing source
NOTE
Make sure to set Frequency Reference for all measurement mode in 5G NR
Signal Analyzer to sync the proper clock source.
shows an example of trigger spectrum measurement.
SPA06MA-O User's Guide
Sync Raster
Offset Range
0 to 11
0 to 5
0 to 5
0 to 2
22145360, R005
SSB Frequency
= Center
Frequency
0
0
0
0
April 2021