10 Tap SSB Periodicity and select the options from 5 ms, 10 ms, 20 ms, 40 ms, 80
ms, 160 ms. The SS burst period (SSB periodicity) is defined by each base
station and network configurable. 20 ms is the default for the initial cell selection.
11 Tap to switch PCI (Physical Cell ID) to Manual or Auto.
12 Tap PCI and input a value from 0 to 1007 using the on-screen keyboard. The PCI
switches to Manual.
13 Tap Symbol Phase Comp and select from the following choices: Auto, Manual,
or Off.
Symbol Phase Compensation is used to compensate for phase differences
between symbols caused by upconversion. Users do not always use the instru-
ment based on RF frequency. In this case, you can set it to Manual or Off.
April 2021
Auto: Setting radio frequency to center frequency
Manual: Setting radio frequency as required
Off: Setting radio frequency to 0
SPA06MA-O User's Guide
22145360, R005
Chapter 7 Using 5G NR Signal Analyzer
Conducting PDSCH measurement