Chapter 2 Using Spectrum Analyzer
Utility
5
Optional: If ACP is selected, set the Main Channel BW and Offset Limit.
a
b
c
d
e
6
Tap the Start button on the right panel of the map to start plotting for the Position
or Time setting.
7
Touch directly on the screen and tap to collect data and plot points on the loaded
map. For the Position setting, you can change the direction of the route with
screen touch. For the Time setting, when you first touch screen a purple triangle
appears on the point and the "Number of n" appears in the bottom left of the map.
The instrument interpolates points in between two enters on a straight dotted line.
The max count is 2560, and the time interval is 0.5 s.
8
Tap the Pause button to pause plotting. The Pause button turns blue and GNSS
cannot be plotted.
9
Tap to switch Plot to Stop to stop plotting.
10 Tap Menu > Sweep > Gated Sweep to set the Gated Sweep On or Off.
11 If you start testing and select Setup > Plot Point > Position, you can undo by
tapping the Start button.
12 If you select the Stop button, the Plot Stop popup window appears.
To set limit
You can set the thresholds for the four different color indicators.
1
Tap Menu > Limit.
2
Tap Excellent and set a value using the on-screen keyboard.
3
Tap Very Good and set a value using the on-screen keyboard.
4
Tap Good and set a value using the on-screen keyboard.
5
Tap Poor and set a value using the on-screen keyboard.
Page 38
Tap Main Channel BW and input a value using the on-screen keyboard.
Tap Offset Limit and tap to switch the Offset to On to display or Off to hide
the offset.
Tap Integration Bandwidth and input a value using the on-screen keyboard.
Tap Offset Frequency and input a value using the on-screen keyboard.
Tap Offset Limit and input a value using the on-screen keyboard.
NOTE
The instrument does not automatically save the collected data. It is recom-
mended that you save the result. If not, you will lose all the collected data.
SPA06MA-O User's Guide