Step 2. Connect the Line- Power Cable
This is a Safety Class I Product (provided with protective earth). The
WA R N I N G
mains plug shall only be inserted in a socket outlet provided with a
protective earth contact. Any interruption of the protective conductor
inside or outside of the instrument is likely to make the instrument
dangerous. Intentional interruption is prohibited.
Always use the three-prong AC power cord supplied with this instrument.
C A U T I O N
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
Do not connect ac power until you have verified the line voltage is correct as
C A U T I O N
described in the following paragraphs. Damage to the equipment could result.
This instrument has autoranging line voltage input. Be sure the supply voltage
C A U T I O N
is within the specified range.
1 Verify that the line power meets the requirements shown in the
following table.
Line Power Requirements
2 Connect the line- power cord to the instrument's rear- panel connector.
Power:
max. 70 W
Voltage:
max: 110 / 115 / 230 / 240 V~
Frequency
50 / 60 Hz
Getting Started

Step 2. Connect the Line-Power Cable

1-5