Installation and Maintenance

AC Line Power Supply Requirements

AC Line Power Supply
Requirements

Line Power Requirements

The Agilent 8163A/B Lightwave Multimeter System complies with
overvoltage category II and can operate from the single-phase AC
power source that supplies between 100 V and 240 V at a frequency in
the range 50 to 60 Hz.The maximum power consumption is 120 VA
with all options installed.
The Agilent 8164A/B Lightwave Measurement System complies with
overvoltage category II and can operate from any single-phase AC
power source that supplies between 100 V and 240 V at a frequency in
the range from 50 to 60 Hz. The maximum power consumption is 270
VA with all options installed.
The Agilent 8166A/B Lightwave Multichannel System complies with
overvoltage category II and can operate from any single-phase AC
power source that supplies between 100 V and 240 V at a frequency in
the range from 50 to 60 Hz. The maximum power consumption is 450
VA with all options installed.

Line Power Cable

In accordance with international safety standards, the instrument has
a three-wire power cable. When connected to an appropriate AC
power receptacle, this cable earths the instrument cabinet. For the
part number of the power cable for your countryand instrument see
"Accessories" on page 297.
.