Status:questionable:calibration Condition Register Bits - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Use Status Registers to Determine the State of Analyzer Events and Conditions
Decimal
Bit
Value
4
16
5
32
6–8
–––
9–14
–––
15
32768
STATus:QUEStionable:CALibration Condition
Register Bits
Decimal
Bit
Value
0, 1
0, 2
2
4
3
8
4
16
5
32
6
64
7
128
Chapter 2
Description
Synth Unlocked A 1 in this bit position indicates
that the analyzer synthesizer is unlocked.
Invalid BW A 1 in this bit position indicates that an
invalid bandwidth setting has been requested.
Reserved These bits are not used by the analyzer, but
are for future use with other Agilent products.
Unused These bits are always set to 0.
Always Zero (0) This bit is always set to 0.
Description
Reserved These bits are not used by the analyzer, but
are for future use with other Agilent products.
TG Align Failure A 1 in this bit position indicates
that a failure has occurred while trying to align the
tracking generator (TG).
RF Align Failure A 1 in this bit position indicates
that a failure has occurred while trying to align the
RF section.
IF Align Failure A 1 in this bit position indicates
that a failure has occurred while trying to align the IF
section.
LO Align Failure A 1 in this bit position indicates
that a failure has occurred while trying to align the
local oscillator (LO).
ADC Align Failure A 1 in this bit position indicates
that a failure has occurred while trying to align the
analog-to-digital converter (ADC).
FM Demod Align Failure A 1 in this bit position
indicates that a failure has occurred while trying to
align the FM demodulation circuitry. (Agilent ESA
models E4401B, E4402B, E4404B, E4405B, and
E4407B Option BAA, FM Demodulation, only).
Status Registers
2-23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents