Standard Event Status Group

These registers are programmed by Common commands. The group consists of an Event and Enable
register.The Standard Event event register latches events relating to communication status. It is a
read-only register that is cleared when read. The Standard Event enable register functions similarly
to the enable registers of the Operation and Questionable status groups. Refer to
a description of each register. The following table describes the bit assignments.
Bit
Bit Name
0
Operation
Complete
1
not used
2
Query Error
3
Device-Specific
Error
4
Execution Error
5
Command
6
not used
7
Power On

Status Byte Register

This register summarizes the information from all other status groups as defined in the IEEE 488.2
Standard Digital Interface for Programmable Instrumentation. The following table describes the bit
assignments.
Bit
Bit Name
0
not used
1
not used
2
Error Queue
3
Questionable Status
Summary
4
Message
Available
5
Event Status
Summary
Keysight N6705C Operating and Service Guide
Decimal
Value
1
All commands before and including *OPC have been executed.
not used
0 is returned
4
The instrument tried to read the output buffer but it was empty, a new command line
was received before a previous query has been read, or both input and output buf-
fers are full.
8
A device-specific error, including a selftest error, calibration error or other device-
specific error occurred.
16
An execution error occurred.
32
A command syntax error occurred.
not used
0 is returned
128
Power has been cycled since the last time the event register was read or cleared.
Decimal
Value
not used
0 is returned
not used
0 is returned
4
One or more errors in the Error Queue. Use SYSTem:ERRor? to read and delete
errors.
8
One or more bits are set in the Questionable Data Register. Bits must be enabled, see
STATus:QUEStionable:ENABle.
16
Data is available in the instrument's output buffer.
32
One or more bits are set in the Standard Event Register. Bits must be enabled, see
*ESE.
7 SCPI Programming Reference
Definition
Error Messages
Error Messages
Error Messages
Definition