Advanced Troubleshooting

Convectron Gauge and Foreline Adapter with Hose

Convectron Gauge and Foreline Adapter with Hose
380
TSQ 8000 Evo Hardware Manual
A convectron gauge measures and controls the pressure in the foreline that is connected to the
turbomolecular pump and rough pump. If the foreline pressure gets too high, convectron
gauge decreases the power going to the turbomolecular pump. The convectron gauge and
foreline adapter are connected to the foreline hose, which is connected to the turbomolecular
pump.
To replace a convectron gauge and foreline adapter with hose
Note See the TSQ 8000 Evo Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
System"
on
page
107. You do not have to remove the ion source cartridge.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 406. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
5. Move the GC away from the TSQ 8000 Evo instrument.
"Shutting Down the TSQ 8000 Evo
Screws
Screw on Front
of Instrument
Thermo Scientific