Figure 100. Removing the Collision Cell and Quadrupole 3 from the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
8. Disconnect the Q1 entrance lens wire from the source interface board.
9. Remove the two screws on each side of the Q1 entrance endcap.
Figure 101. Removing Screws from the Quadrupole Rod Endcaps
Q1 Entrance
Lens and Endcap
Screws
10. Remove the Q1 entrance lens endcap and ion guide from the tray.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Q3 Exit Lens
Assembly
Screws
Collision Cell
Assembly
TSQ 8000 Evo Hardware Manual
127