Removing the Analyzer Tray from the TSQ 8000 Evo Instrument
Thermo Scientific
To remove the analyzer tray from the TSQ 8000 Evo instrument
1. Slide the top cover panel toward the back of the instrument and lift it off.
2. Disconnect the electrometer board cable
Figure 72. Disconnecting the Electrometer Board Cable
Electrometer
Board Cable
3. Remove the small and large glass manifold cover.
4. Set the glass covers on a surface where they will not be scratched or contaminated. Protect
the glass and any other manifold components from dust accumulation. Wear cleanroom
gloves when touching the glass. See the TSQ 8000 Evo Spare Parts Guide for information
about ordering cleanroom gloves.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
TSQ 8000 Evo Hardware Manual
109