Thermo Scientific TSQ 8000 Evo Hardware Manual page 137

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Thermo Scientific
Figure 75. Disconnecting the Quad 3 Connector Wires and the Collision Cell Connector Wires
Rear 4-pin
Connector
Collision Cell
Connector Wires
7. Remove the collision gas tube from the collision cell.
Note Pull straight back from the collision cell to prevent damage to the collision gas tube.
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
Figure 76. Opening the Manifold Door
9. Open the manifold door and slide the filament board and wires off the pins on the source
interface board
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Screws
TSQ 8000 Evo Hardware Manual
Quad 3
Wires
Collision Gas
Tube
111