Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning

Reassembling the Ion Source Cartridge

40
TSQ 8000 Evo Hardware Manual
1. Place the ion volume-repeller insulator on the repeller and hold it in place with the
repeller nut. Set it aside for now.
2. Insert the ion cartridge sleeve into the source holder.
Figure 31. Inserting the Sleeve into the Source Holder
3. Align the long tooth of lens 3/RF lens with the notch on the bottom of the sleeve and
drop the lens into the sleeve.
Figure 32. Inserting Lens 3/RF Lens into the Source Sleeve
Long Tooth on Lens 3/RF Lens
Tooth Fits Here