6. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
Figure 407. Removing the Right Side Panel Screws
7. Disconnect the vacuum hose from the adapter on the foreline pump.
8. Remove the rough pump clamp on the turbomolecular pump and the adapter.
Convectron Gauge and Foreline Adapter with Hose
Right Side
Panel Screws
TSQ 8000 Evo Hardware Manual
3
Advanced Troubleshooting
381