Advanced Troubleshooting
Replacing Components of the Analyzer
130
TSQ 8000 Evo Hardware Manual
Figure 105. Replacing the Collision Cell and Quadrupole 3 in the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
20. Reconnect the three lens wires attached to the lenses between the quadrupoles.
21. Reconnect the Q1 entrance lens wire to the source interface board.
22. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray in the TSQ 8000 Evo Instrument"
replace the tray in the TSQ instrument.
23. Restart your system by following the instructions in
Q3 Exit Lens
Assembly
Screws
on
page 118
"Restarting the System"
on
Thermo Scientific
to
page
120.