Advanced Troubleshooting
Replacing Components of the Ion Source
166
TSQ 8000 Evo Hardware Manual
4. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block, but leave it attached to the tray.
Figure 152. Disconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
5. Use a T20 Torxhead screwdriver to remove the two screws on the back of source block.
Figure 153. Disconnecting the Source Block to the Analyzer Tray
Screws
Note Some of the pieces may detach as you loosen the screws. Do not let them fall.