17. Flip door over and insert and tighten the vent valve knob and o-ring into the front
manifold plate.
Figure 242. Reattaching the Vent Valve Knob and O-Ring
Vent Valve
O-ring
Vent
Valve
Knob
18. Reattach the three screws and metal block to the front manifold plate.
Figure 243. Reattaching the Metal Block
Screws
3
Advanced Troubleshooting
Replacing Components of the Manifold