3
Advanced Troubleshooting
Replacing Components of the Analyzer
110
TSQ 8000 Evo Hardware Manual
Figure 73. Removing the Small and Large Glass Manifold Covers
5. Disconnect the four wires that go to the ion guide and quadrupole 1. These wires are
attached to the front 4-pin feedthrough in the vacuum manifold.
Figure 74. Disconnecting the Ion Guide Connector Wires and the Quad 1 Connector Wires
Front 4-pin
Connector
Ion Guide Wires
6. Disconnect the four wires that go to the collision cell and quadrupole 3. These wires are
attached to the rear 4-pin feedthrough in the vacuum manifold.
Large Glass
Manifold Cover
Small Glass
Manifold Cover
Quad 1
Wires