3
Advanced Troubleshooting
Replacing the Boards in the TSQ 8000 Evo Instrument
296
TSQ 8000 Evo Hardware Manual
8. Remove the controller interface board and PC communication board from the
instrument.
Figure 319. Removing the Controller Interface Board Support Bracket
9. If you need to replace the support bracket, use a T10 Torxhead screwdriver to remove the
seven M3 screws securing the controller interface board to the support bracket.
10. Replace the bracket and reattach it to the controller interface board.
Controller
Interface Board
Support Bracket
Screws
PC Communication Board
Thermo Scientific