Thermo Scientific TSQ 8000 Evo Hardware Manual page 244

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Manifold
218
TSQ 8000 Evo Hardware Manual
12. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
tighten the screw holding the lens driver board in place on the chassis.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
Figure 229. Replacing the Lens Driver Board
13. Replace the left hand front panel and tighten the two screws.
14. Restart your system by following the instructions in
Lens Driver Board
Screw
"Restarting the System"
on
page
120.