Q3 Exit Lens Assembly

Thermo Scientific
To replace the Q3 exit lens assembly
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 8000 Evo Instrument"
3. Turn the tray to access the bottom and remove the wire to the Q3 exit lens.
4. Remove the two screws on each side of the Q3 exit lens assembly and slide it off the back
of the tray while holding the Q3 quad assembly in place.
Figure 115. Removing the Q3 Exit Lens Assembly from the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
5. Attach a new Q3 exit lens assembly and reattach the two screws.
Tip You will get better signal if you gently push the Q3 exit lens assembly forward while
tightening the screws.
6. Reattach the two screws on each side of the quadrupole rod endcaps.
Tip Look through the endcaps to be sure the screws are aligned properly.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
"Shutting Down the TSQ 8000 Evo
on
page
Q3 Exit Lens
Assembly
TSQ 8000 Evo Hardware Manual
109.
Screws
139