7. Use a 2.5 mm nut driver to remove the standoffs on the rod driver board.
Figure 288. Disconnecting the Rod Driver Board from the RF Board
Screws
8. Incrementally, remove the 40-pin connector on the rod driver board from the front RF
board. and the 7-pin connector on the rod driver board away from the front RF board.
Figure 289. Disconnecting the Rod Driver Board 40-Pin Connector
3
Replacing the Boards in the TSQ 8000 Evo Instrument
Standoffs
Rod Driver Board
40-Pin Connector
TSQ 8000 Evo Hardware Manual